How do you sterilize instruments in Cidex?
Immerse clean, dry instru- ments completely in the CIDEX OPA Solution. Ensure all instruments are completely submersed in CIDEX OPA Solution, and if applicable, fill all lumens. Cover the CIDEXSolution Tray with a secure lid. Soak instruments for 12 minutes at 20ºC to achieve high level disinfection.
How does Cidex sterilization work?
CIDEX® OPA Solution is a high-level disinfectant for reprocessing reusable heat-sensitive semi-critical medical devices such as endoscopic, respiratory therapy, and anesthesia equipment. This product is not a sterilant. Ensure adequate ventilation and wear the appropriate PPE.
What is the time required for sterilization of articles in Cidex?
10 hours
The chemical is considered bactericidal in 10 minutes, destroying all bacteria, including Myobacterium, tuberculosis, Pseudomonas aeruginosa, and viruses. It is sporicidal in 10 hours and, therefore, considered a sterilizing agent after use for 10 hours (Johnson et al, 1982).
Who manufactures Cidex Opa?
Advanced Sterilization Products
CIDEX products are manufactured by Advanced Sterilization Products, a division of Ethicon (a Johnson & Johnson company). CIDEX OPA is an effective medical instrument disinfectant and it is not to be used on skin.
What is Cidex made of?
CIDEX® OPA Solution is a clear, pale blue liquid with a pH of 7.5. It contains 0.55% ortho-phthalaldehyde in an aqueous base containing buffers,chelating agents and a corrosion inhibitor.
What color is cidex?
Light blue
Liquid color: Light blue. Use: Manufacturer recommends wearing eye protection and chemical-resistant gloves. Solubility: Water-soluble.
What type of disinfectant is cidex?
DESCRIPTION: CIDEX® OPA Solution is a High Level Disinfectant (HLD) for use in re-processing heat sensitive medical devices. CIDEX OPA Solution is the first new HLD available in the last thirty years with the broad materials compatibility of glutaraldehyde.
Why is 70% alcohol used for sterilization not 100?
A 70% solution of alcohol takes more time in evaporation from the surface, increasing the contact time. Therefore, 70% isopropyl alcohol fulfills both requirements. 100% isopropyl alcohol coagulates the protein instantly creating a protein layer that protects the remaining protein from further coagulation.
Why is 100 ethanol not used for sterilization?
If we use 100% alcohol it will evaporate quickly so the time required for alcohol to show its effect will be reduced, similarly for diluted one the contact time will be increased but the minimum toxic concentration which is required for cell death will not be achieved.
Why is 70% ethanol a better disinfectant than 100% ethanol?
70% denatured alcohol penetrate the cell wall more completely which permeates the entire cell, coagulates all proteins, and therefore the microorganism dies. Extra water content slows evaporation, therefore increasing surface contact time and enhancing effectiveness.
How long does Cidex disinfection solution last?
1 It achieves high level disinfection in 20 minutes at 20°C making it highly effective. 2 It is a long lasting solution which is reusable for up to 14 days and can be monitored with Cidex Solution Test Strips. 3 It is compatible with and can be used on a wide range of instruments including the Welch Allyn 60813.
What is the MEC of Cidex ® OPA solution?
CIDEX ® OPA Solution Test Strips are developed exclusively for monitoring the minimum effective concentration (MEC) of CIDEX ® OPA Solution. The CIDEX ® OPA Solution Test Strips provide a rapid and easy measure of the minimum effective concentration (MEC) to enable optimal reuse of CIDEX ® OPA Solution for high-level disinfection.
What is Cidex solution used for?
Cidex solution has been one of the most trusted sterilizing agents in the medical field. It is frequently used for cleaning and sterilizing medical tools especially those used in the operating room. And medical personnel, whether you’re a doctor, nurse or hospital staff, have the proper knowledge on the use of this chemical disinfectant.
How do you sterilize with Cidex?
1 Clean the items to be sterilized before sterilizing to ensure effectiveness. 2 The recommended soaking time is within 10-45 minutes. 3 Completely immerse the items to be disinfected in the solution. 4 Rinse the equipment in sterile water after soaking in Cidex and protect from re-contamination until used.